Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 18002 |
| Brand Name: | Baxter MiniCap with Povidone-Iodine Solution |
| Model: | Product Code: 5C4466P |
| Affected: | All non expired lots |
| Software version: | |
| Recalling Organisation: | Baxter Healthcare Ltd 33 Vestey Drive Mt Wellington AUCKLAND |
| Contact Information: | 0800 229 837 |
| Manufacturer: | Baxter Healthcare Corporation SA |
| Issue: | Baxter received complaints indicating that the sponge of the MiniCap was fully separated from the cap, partially protruding from the cap, or missing. Use of MiniCaps with sponges fully separated or missing from the caps may compromise the ability of the MiniCap to provide a sterile barrier protection at the end of the transfer when the transfer set is not connected to the patient line of the automated peritoneal dialysis (APD) cassette or continuous ambulatory peritoneal dialysis (CAPD) twin bag set-ups. This may increase the risk of peritonitis. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 11/2/2015 |