Detail of Recall

Type of Product:Device
Medsafe Reference:17989
Brand Name:Zimmer Femoral Heads and Taper
Model:32 models
Affected:Product distributed between May-2005 through Nov-2014
Software version:
Recalling Organisation:Zimmer Biomet New Zealand Company
210 Khyber Pass Road
Grafton
Auckland
Contact Information:09 925 5200
Manufacturer:Zimmer GmbH
Issue:Zimmer identified a subset of products that were manufactured prior to mid-2012 and packaged in a specific configuration that failed a packaging test due to potential compromise of the inner sterile tray. This packaging configuration consists of a double sterile barrier. There are no known incidents where the outer tray was compromised; therefore the device remains sterile when delivered during surgery. This recall is not related to sterility concerns, but rather the removal of any unconsumed product given that the packaging may not have the double sterile barrier as intended.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/1/2015