Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:17987
Brand Name:Siemens syngo.Plaza Imaging System
Model:Catalogue Number: 10863171 & 10592457
Affected:
Software version:All versions
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Issue:In case of a system crash images may not be written to the hard disk and there may be inconsistencies in the database. When a system crash occurs (e.g. blue screen, power outage), images may not be written from cache to the hard disk and are lost. In such a case an existing file name identifier might be re-used after restart of the system. Now two references to the same image file on STS exist in the database of syngo.plaza.
Manufacturer has now released another version of the software as a final correction to this this action (SY099/14/s) . THis software update will fix any inconsistencies in the archived data that may have occurred before software update VB10A_HF06 was installed.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/1/2016