Detail of Recall

Type of Product:Device
Medsafe Reference:17947
Brand Name:HeartWare Controller
Model:Product Codes: 1400, 1401XX
Affected:Serial Number Range: CON000001 through CON005472
Software version:
Recalling Organisation:Health New Zealand - Te Whatu Ora
9 Ruskin Street
Parnell
Auckland 1052
Contact Information:
Manufacturer:HeartWare International Inc
Issue:A selected range of serial numbers of HeartWare Controllers exhibit a higher susceptibility to electrostatic discharge (ESD) than new Controllers. In 2013 HeartWare advised of the potential for this issue. Since then the manufacturer has received reports of one additional death and one additional serious injury in which ESD may have caused or contributed to data corruption in older Controllers resulting in the loss of commutation to pump motors, which ultimately led to pump stops.
Recall action type:Recall
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/1/2015