Type of Product:Device
Medsafe Reference:17924
Brand Name:Stryker Triathlon Tibial Alignment Ankle Clamp EM
Model:Catalogue Number: 65412609. Affected Kits: 65418002-T, 65418030-T, TRIAR36PREP-T, WKSHP65418002-T & SHAPECONV-T
Affected:Lot Numbers: Lots prior to X9L06
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Stryker Howmedica Osteonics
Issue:Stryker Orthopaedics carried out a recall for product correction in July 2013 due to complaints associated with cracks or fracture of the Plastic Triathlon Tibial Alignment Ankle Clamp (old design). This issue may lead to potential complications during Surgery in the event a replacement sterile ankle clamp is required mid-procedure, which may create a delay exceeding 30 minutes in duration. Stryker will organise the swap out of any old plastic clamps for the new metal clamps.
Recall action type:Recall
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:27/2/2015