Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:17922
Brand Name:Radiometer ABL90Flex analyser
Model:
Affected:
Software version:
Recalling Organisation:Radiometer Pacific
Suite 1
1/30 Benedicts Street
Eden Terrace
Auckland 1010
Contact Information:0800 723 722
Manufacturer:Radiometer Medical ApS
Issue:Manufacturer has become aware of a potential significant clinical issue with patient results if the inlet is left open after aspiration of the patient sample for more than 30 seconds.
Note that if the inlet is accidentally left open after processing a patient sample the ABL90 analyser sounds a warning bell, places a warning message on the screen and also initiates a voice message continuously repeating "Please close the inlet".
The problem can occur after the following sequence of actions:
1.The operator aspirates patient sample A and forgets to close the inlet
2.The inlet is left open for more than 30 seconds before it is closed
3.An operator aspirates patient sample B
In this case the results for patient sample B are affected by this issue
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/12/2014