Detail of Recall

Type of Product:Device
Medsafe Reference:17913
Brand Name:Siemens Digital Linear Accelerators of type MEVATRON, ARTISTE , PRIMUS, ONCOR and ARTISTE
Model:08139789, 05864472, 05857912, 07360717, 05857920, 04504200, 01940035, 05500371, 09822685, 09411588, 09822693, 08515520, 01940753, 01924500, 09401506, 09401407, 09401316, 10941746 and 09401654
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Issue:Manufacturer has received reports of one electric board (G41) located inside of the linac gantry showing signs of over heating. After a detailed analysis the root cause was found to be one of the components that belong to this board (Head Driver PCB). The manufacturer has redesigned the Head Driver PCB to further mitigate the risk of local over heating on the G41 motherboard. The present corrective action comprises the replacement of the Head Driver PCB on afected systems.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/12/2014