Detail of Recall

Type of Product:Device
Medsafe Reference:17909
Brand Name:Monaco Image Statistics Tool
Model:Monaco
Affected:Rel 3.30 and higher
Software version:
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Elekta Inc
Brand Name:Monaco Image Statistics Tool
Model:Monaco
Affected:
Software version:Rel 3.30 and higher
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Elekta Inc
Issue:When a CT phantom is scanned and the images are transferred to Monaco, it is possible to sample the densities using the Image Statistics tool in Monaco. In the case where CT slice thickness is not the same as CT slice spacing the Image Statistics tool will report an incorrect HU. If the user bases their CT to ED file on this incorrect density information, dose calculation errors will result. The magnitude of the dose calculation error will be dependent on the magnitude of the HU error.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/12/2014