Detail of Recall

Type of Product:Device
Medsafe Reference:17899
Brand Name:St Jude Medical Merlin @ Home
Model:Model: EX1150
Affected:All units
Software version:
Recalling Organisation:Abbott Medical New Zealand Limited
4 Pacific Rise
Mt Wellingtion
Auckland 1640
Contact Information:0800 756 269
Manufacturer:St Jude Medical (USA)
Issue:Low incidence of back up operation in some implanted St. Jude Medical devices with Radio Frequency (RF) capability. This may occur as a result of a Merlin@home transmitter initiating an implanted device software reset.
This issue can only occur when the patient is being actively monitored by a Merlin@home RF bedside transmitter.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/12/2014