Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:17880
Brand Name:Alere Actim Partus
Model:Catalogue Numbers: 31930ETAL, 31931ETAL, 31931ETAC
Affected:Multiple lot numbers
Software version:
Recalling Organisation:Abbott Laboratories (NZ) Ltd - Rapid Diagnostics Divison
Ground Floor, Bldg. D
4 Pacific Rise
Mount Wellington
AUCKLAND 1060
Contact Information:0800 466 951
Manufacturer:Medix Biochemica Oy Ab
Issue:Certain batches of Actim Partus dipsticks have shown increased non-specific binding during late stages of the 24 month shelf-life of the product. This issue increases the risk of false-positives results. The issue does not affect negative results which are still valid.
Recall action type:Recall
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/12/2014
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /