Type of Product:Device
Medsafe Reference:17855
Brand Name:GE Proteus XR X-ray system
Model:
Affected:Devices with wall stand model number 600-0301 manf before Jan-2011
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Hualun Medical Systems Co Ltd
Issue:GE Healthcare has recently become aware of a bucky device falling from the wall stand as a result of the breakage of its supporting cables. A fall of a bucky while the system is in use could result in an injury to a patient or operator. There were no injuries reported as a result of this issue.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/12/2014