Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 17852 |
| Brand Name: | Alere INRatio & INRatio 2 PT/INR Monitoring System |
| Model: | All models |
| Affected: | All lots |
| Software version: | |
| Recalling Organisation: | Abbott Laboratories (NZ) Ltd - Rapid Diagnostics Divison Ground Floor, Bldg. D 4 Pacific Rise Mount Wellington AUCKLAND 1060 |
| Contact Information: | 0800 466 951 |
| Manufacturer: | Alere San Diego Inc |
| Issue: | The INRatio® and INRatio®2 PT/INR Monitor system should not be used on patients with any of the following conditions: * Anemia of any type with hematocrit less than 30% * Any conditions associated with elevated fibrinogen levels including: - Acute inflammatory conditions (examples may include acute viral or bacterial infections such as pneumonia or influenza) - Chronic inflammatory conditions (examples may include rheumatoid arthritis, Crohn's disease, ulcerative colitis, infectious liver diseases such as hepatitis, or inflammatory kidney diseases such as diabetic nephropathy and glomerulonephritis) - Severe infection (e.g., sepsis) - Chronically elevated fibrinogen for any reason - Hospitalized or advanced stage cancer or end stage renal disease patients requiring hemodialysis * Any bleeding or unusual bruising, clinically observed or reported by the patient |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Public Recall |
| Date Commenced: | 24/12/2014 |