Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:17849
Brand Name:Philips Sync Cables
Model:Models: M1783A and M5526A
Affected:All Philips sync cables
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:When a Philips monitor/defibrillator is receiving an ECG signal from an auxiliary bedside monitor via a sync cable, the following can occur if the monitor/defibrillator experiences interference from electrical fast transients (EFTs) while connected to AC power:
* On the HeartStart MRx and HeartStart XL, EFT noise can be misinterpreted as an R-wave
* On the HeartStart XL+, EFT noise can disable ECG monitoring, and potentially interrupt demand mode pacing.

It is contrary to the instructions in the XL+ Instructions for Use (IFU) to perform demand mode pacing while using the sync cable to provide the ECG signal from a beside monitor.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/12/2014