Detail of Recall

Type of Product:Device
Medsafe Reference:17839
Brand Name:Philips HeartStart MRx Monitor/Defibrillator
Model:Models: M3535A & M3536A
Affected:Selected serial numbers
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:The MRx could contain incorrect internal software settings, causing the following two issues:
1. The device will perform the weekly automated tests hourly, which could cause the therapy capacitors to degrade sooner than intended.
2. While connected to AC or DC power and with no battery installed or the battery installed has a charge level of less than 10%, the Ready for Use (RFU) indicator will not provide the expected low battery indication (flashing red X with audible chirp). Instead, the RFU will show a flashing black hourglass, indicating that sufficient battery power is available for device operation.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/12/2014