Type of Product:Device
Medsafe Reference:17828
Brand Name:Carestream DRX Revolution X-Ray System
Model:
Affected:Manufactured prior to 15-Oct-2013
Software version:
Recalling Organisation:Carestream Health New Zealand Ltd
c/- Quigg Partners
Level 7, 28 Brandon St
Wellington
Contact Information:021 977 215
Manufacturer:Carestream Health Inc (HCIS)
Issue:The manufacturer has received a report that an overseas service engineer observed a loose boom arm during routine maintenance on a DRX Revolution Mobile X-Ray System as one of three bolts that secure the boom to the carriage had fractured and the other two bolts were loose causing the boom arm to become loose. No other reports of this type have been received since the device was introduced into commercial distribution in 2009.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:4/12/2014