Detail of Recall

Type of Product:Device
Medsafe Reference:17809
Brand Name:Rusch Bronchopart Tube Set Left
Model:Product Code: 116100-000260
Affected:29 lots
Software version:
Recalling Organisation:Ebos Group Ltd
108 Wrights Road
Addington
Christchurch
Contact Information:0800 18 17 16
Manufacturer:Rusch Inc
Issue:Teleflex are recalling specific products following receipt of customer complaints reporting that the cobb connector detached from the main connector prior to use. This cobb connector detachment should be readily identified by the user during set-up and prior to use of the device. If during use, cobb connector separation occurs and presents as a leak in the breathing circuit and if it is unnoticed, then there is a remote probability that the patient may need to be re-intubated due to the risk of respiratory distress and hypoxia.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/1/2015