| Type of Product: | Device |
| Medsafe Reference: | 17809 |
| Brand Name: | Rusch Bronchopart Tube Set Left |
| Model: | Product Code: 116100-000260 |
| Affected: | 29 lots |
| Software version: | |
| Recalling Organisation: | Ebos Group Ltd 108 Wrights Road Addington Christchurch |
| Contact Information: | 0800 18 17 16 |
| Manufacturer: | Rusch Inc |
| Issue: | Teleflex are recalling specific products following receipt of customer complaints reporting that the cobb connector detached from the main connector prior to use. This cobb connector detachment should be readily identified by the user during set-up and prior to use of the device. If during use, cobb connector separation occurs and presents as a leak in the breathing circuit and if it is unnoticed, then there is a remote probability that the patient may need to be re-intubated due to the risk of respiratory distress and hypoxia. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 12/1/2015 |