Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:17784
Brand Name:Halyard KimVent Subglottic Suctioning Endotracheal Tubes
Model:Product Code: 13222
Affected:Lot Number: AA4062V04
Software version:
Recalling Organisation:Pharmacy Retailing (NZ) Ltd t/a Healthcare Logistics
54 Carbine Road
Mt Wellington
AUCKLAND
Contact Information:09 969 0705
Manufacturer:Kimberly-Clark Corp
Issue:Halyard Health's Quality Assurance processes have identified that the inflation line of certain KimVent* Microcuff* Subglottic Suctioning Endotracheal Tubes may detach from the endotracheal tube during use (i.e., if pulled, tugged, or excess tube/patient movement occurs). Detachment of the inflation line leads to gradual deflation of the endotracheal tube cuff potentially leading to respiratory circuit air leak between the cuff and tracheal wall.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/12/2014
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /