Detail of Recall

Type of Product:Device
Medsafe Reference:17752
Brand Name:GE Revolution XR/d X-ray systems
Model:
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Medical Systems LLC
Brand Name:GE Proteus XR/a X-ray systems
Model:
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Hualun Medical Systems Co Ltd
Issue:The Proteus XR/a and Revolution XR/d X-ray service manuals have been updated to further ensure proper re-installation. Reinstallation errors may result in a fall of a collimator while the system is in use, resulting in an injury to a patient or operator. There was a reported incident of a collimator fall which caused a serious patient injury.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:25/11/2014