Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 17752 |
| Brand Name: | GE Revolution XR/d X-ray systems |
| Model: | |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Medical Systems LLC |
| Brand Name: | GE Proteus XR/a X-ray systems |
| Model: | |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Hualun Medical Systems Co Ltd |
| Issue: | The Proteus XR/a and Revolution XR/d X-ray service manuals have been updated to further ensure proper re-installation. Reinstallation errors may result in a fall of a collimator while the system is in use, resulting in an injury to a patient or operator. There was a reported incident of a collimator fall which caused a serious patient injury. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 25/11/2014 |