Type of Product:Device
Medsafe Reference:17745
Brand Name:Aircraft Medical McGrath MAC Video Laryngoscope Blade Size#2
Model:350-017-000
Affected:Lot numbers: 3121202, 13122301, 13042401, 14030501, 14031814 and 14051616
Software version:
Recalling Organisation:Covidien NZ Ltd
Level 3, Building 5, Central Park Corporate Centre
666 Great South Road
Penrose
AUCKLAND 1051
Contact Information:0508 489 264
Manufacturer:Aircraft Medical Ltd
Issue:When a blade from an affected Lot is fited to certain MAC Video Laryngscopes it may be difficult to get the blade clip to engage onto the tooth on the handle. This increases the possibility of blade becoming disconnected during intubation specifically when the device is being retracted from the mouth of the patient.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/11/2014