Detail of Recall

Type of Product:Device
Medsafe Reference:17677
Brand Name:Radiometer 942-073, D8088 and D8089 cCrea membrane
Model:942-073, D8088 and D8089
Affected:cCrea membrane Lot 557 onwards
Software version:
Recalling Organisation:Radiometer Pacific
Suite 1
1/30 Benedicts Street
Eden Terrace
Auckland 1010
Contact Information:0800 723 722
Manufacturer:Radiometer Medical ApS
Issue:RADIOMETER has become aware that the cCrea results for QC decreases to below the lower acceptance limit within the lifetime of the Crea membranes. This has been experienced on membranes down to five days after installation of new membrane units.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/11/2014