Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 17677 |
| Brand Name: | Radiometer 942-073, D8088 and D8089 cCrea membrane |
| Model: | 942-073, D8088 and D8089 |
| Affected: | cCrea membrane Lot 557 onwards |
| Software version: | |
| Recalling Organisation: | Radiometer Pacific Suite 1 1/30 Benedicts Street Eden Terrace Auckland 1010 |
| Contact Information: | 0800 723 722 |
| Manufacturer: | Radiometer Medical ApS |
| Issue: | RADIOMETER has become aware that the cCrea results for QC decreases to below the lower acceptance limit within the lifetime of the Crea membranes. This has been experienced on membranes down to five days after installation of new membrane units. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 12/11/2014 |