Detail of Recall

Type of Product:Device
Medsafe Reference:17675
Brand Name:Siemens Advia Liquid Specific Protein (LSP) Calibrator
Model:Catalogue Number: 07711199
Affected:Lot Number: 303574
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Siemens Healthcare Diagnostics has confirmed two issues with the Siemens Liquid Specific Protein (LSP) calibrator lot 303574:
Issue 1: Siemens has observed an increased frequency of Max Fit Deviation flags using the Liquid Specific Protein (LSP) calibrator kit lot 303574 (vial lots 122IT - 127IT) to calibrate the Alpha-1-Antitrypsin (AAT) assay.
Issue 2: Siemens has confirmed that a preliminary version of the Liquid Specific Protein (LSP) calibrator lot-specific Instructions For Use (IFU) for calibrator kit lot 303574 resided on the Siemens Document Library from December 10, 2013 until April 7, 2014. The LSP calibrator IFU version that resided on the Siemens Document Library during this time contained preliminary calibrator values that differed slightly from the final calibrator values for the following ADVIA Chemistry Systems assays: AAG, AAT, ASO_2, C3, C4, FRT, HAPT, IGA_2, IGG_2, IGM_2, PREALB, RF and TRF.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/11/2014