Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 17661 |
| Brand Name: | Philips MammoDiagnost DR System with stereotactic biopsy device |
| Model: | |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems DMC GmbH |
| Issue: | Philips has identified that the pin implemented in the safety switch may not put enough pressure on the safety circuit to prohibit movement. The needle positioning device may move even with the safety switch engaged. If this occurs while the needle is inserted, it could lead to an unintended patient injury. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 10/11/2014 |