Type of Product:Device
Medsafe Reference:17661
Brand Name:Philips MammoDiagnost DR System with stereotactic biopsy device
Model:
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems DMC GmbH
Issue:Philips has identified that the pin implemented in the safety switch may not put enough pressure on the safety circuit to prohibit movement. The needle positioning device may move even with the safety switch engaged. If this occurs while the needle is inserted, it could lead to an unintended patient injury.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/11/2014