Detail of Recall

Type of Product:Device
Medsafe Reference:17647
Brand Name:Bio-Rad IH-1000 software
Model:Catalogue Number: 001000V
Affected:
Software version:All versions
Recalling Organisation:Bio-Rad New Zealand
Level 30
48 Shortland Street
AUCKLAND 1010
Contact Information:0508 805 500
Manufacturer:Bio-Rad Laboratories Inc
Issue:Bio-Rad has confirmed that under specific conditions, the IH-1000 software may lose the link sample/"test in progress". This loss, might, in the worst case, lead to a mismatch between the test result and the sample.
This issue is linked to the sample management when using the "emergency* button" and might occur ONLY in the following conditions:
1. A Not Urgent rack (green label) is inserted with more than one sample AND ;
2. ALL the samples on this rack are manually selected as "Emergency" by clicking on the "Emergency* button";
Then once the first emergency sample is processed, the instrument ejects the rack before all processing is completed.
3. If a rack is then reloaded with new samples in the exact same positions and reinserted in the same slot, the link sample/"test in progress" is lost.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/11/2014