Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 17644 |
| Brand Name: | Siemens Dimension Vista 500 & 1500 Intelligent Lab System |
| Model: | Catalogue number: 10488224 and 10444802 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare Diagnostics Inc |
| Issue: | Siemens has received customer complaints of discrepant flagged and/or unflagged patient and QC results on Dimension Vista® Intelligent Lab Systems. Siemens Healthcare Diagnostics has confirmed that reagent probe performance may decline more quickly than anticipated and have an effect on assays with similar formulations when processed sequentially on the same reagent server. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 6/11/2014 |