Brand Name:Siemens Dimension Vista 500 & 1500 Intelligent Lab System
Model:Catalogue number: 10488224 and 10444802
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Siemens has received customer complaints of discrepant flagged and/or unflagged patient and QC results on Dimension Vista® Intelligent Lab Systems. Siemens Healthcare Diagnostics has confirmed that reagent probe performance may decline more quickly than anticipated and have an effect on assays with similar formulations when processed sequentially on the same reagent server.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/11/2014