Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:17640
Brand Name:FPH IW900-series Infant Radiant Warmers
Model:Model Numbers: IW910xxx, IW920xxx, IW931xxx, IW932xxx, IW933xxx, IW934xxx, IW951xxx, IW952xxx, IW953xxx, IW954xxx, IW980xxx, IW990xxx
Affected:Date of manufacture: 16- Nov-2007 to 10-Aug-2009
Software version:
Recalling Organisation:Fisher & Paykel Healthcare Ltd
15 Maurice Paykel Place
East Tamaki
AUCKLAND 2013
Contact Information:0800 503 553
Manufacturer:Fisher & Paykel Healthcare Ltd
Issue:The manufacturer is recalling selected infant warmers because it has had two confirmed reports of breakage of the nut that secures the heater head in place. Breakage of the nut can cause the heater head to become partially detached and swing towards the bassinet. The heater head did not contact the patient in either case, however, there is the potential that a heater head could contact a patient situated on the bassinet and cause a serious injury.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/11/2014