Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 17640 |
| Brand Name: | FPH IW900-series Infant Radiant Warmers |
| Model: | Model Numbers: IW910xxx, IW920xxx, IW931xxx, IW932xxx, IW933xxx, IW934xxx, IW951xxx, IW952xxx, IW953xxx, IW954xxx, IW980xxx, IW990xxx |
| Affected: | Date of manufacture: 16- Nov-2007 to 10-Aug-2009 |
| Software version: | |
| Recalling Organisation: | Fisher & Paykel Healthcare Ltd 15 Maurice Paykel Place East Tamaki AUCKLAND 2013 |
| Contact Information: | 0800 503 553 |
| Manufacturer: | Fisher & Paykel Healthcare Ltd |
| Issue: | The manufacturer is recalling selected infant warmers because it has had two confirmed reports of breakage of the nut that secures the heater head in place. Breakage of the nut can cause the heater head to become partially detached and swing towards the bassinet. The heater head did not contact the patient in either case, however, there is the potential that a heater head could contact a patient situated on the bassinet and cause a serious injury. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 6/11/2014 |