Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 17639 |
| Brand Name: | DePuy Synthes Attune Intuition Distal Femoral Jig |
| Model: | Part numbers: 254400521 and 254400520 |
| Affected: | All lots |
| Software version: | |
| Recalling Organisation: | Johnson & Johnson (NZ) Limited 13a Gabador Place Mt Wellington AUCKLAND 1060 |
| Contact Information: | 0800 922 638 |
| Manufacturer: | DePuy Ireland |
| Issue: | This correction is being issued because if the metal pin bushing is overloaded, there is the potential for partial or full displacement of the metal pin bushing along with fracture of the plastic around the metal pin bushing. If the plastic around the metal pin bushing fractures, there is the potential for fragments of the fractured plastic to be left in the patient. |
| Recall action type: | Recall |
| Recall action: | Product to be exchanged |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 14/11/2014 |