Type of Product:Device
Medsafe Reference:17594
Brand Name:Philips IntelliVue Patient Monitors
Model:Models: MP5 (M8105A), MP5SC (M8105AS), MP5T (M8105AT), MP5 Upgrades (M8105AU, 866327)
Affected:
Software version:Software revisions J.21.03, J.21.19
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medizin Systeme Boblingen GmbH
Brand Name:Philips Avalon Fetal Monitors
Model:Models: FM20 (M2702A), FM30 (M2703A), FM50 (M2705A)
Affected:
Software version:Software revision J.30.58
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medizin Systeme Boblingen GmbH
Issue:In time-synchronized automatic/sequence mode, the non-invasive blood pressure (NBP) automatic measurement series is stopped inadvertently if the repetition interval is increased manually by the user (e.g. from 5 to 15 minutes) in between two consecutive measurements. If the NBP measurement is stopped, and a patient's blood pressure is not being monitored, a delay in treatment could occur.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/10/2014