Detail of Recall

Type of Product:Device
Medsafe Reference:17574
Brand Name:Medtronic StealthStation AxiEM or Fusion EM Surgical Navigation System
Model:
Affected:All units
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Navigation Inc
Issue:Through testing conducted with these two products by Berchtold and Medtronic, when the lights are positioned according to Berchtold's Instructions for Use - at a distance at least one meter from the patient - there is no interference. When the lights are positioned closer to the patient, the resulting electromagnetic interference (EMI) causes the StealthStation® AxiEM™ or Fusion EM system to cease tracking and display "red status", which prevents navigation until the EMI is removed.
Recall action type:Alert
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/10/2014