Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 17559 |
| Brand Name: | Rocket IPC Drainage System |
| Model: | R54400, R54400-00-PP, R54410, R51400, R51400-00-PP, R51410 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Universal Specialities Ltd trading as USL Medical 494 Rosebank Road Avondale AUCKLAND 1026 |
| Contact Information: | 09 829 0595 |
| Manufacturer: | Rocket Medical PLC |
| Issue: | A small number of drainage bottles have appeared to malfunction as the blue connector tube has not opened once the white clip has been moved. The non-opening of the blue tube does not allow the vacuum contained in the bottle to initiate drainage of fluid from the patient. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 31/10/2014 |