Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 17540 |
| Brand Name: | Stryker SmartLife Large Asceptic Housing |
| Model: | 7126120000 |
| Affected: | 13027 to 14093 |
| Software version: | |
| Recalling Organisation: | Stryker New Zealand Ltd 515 Mount Wellington Highway Mount Wellington AUCKLAND 1060 |
| Contact Information: | 0800 787 9287 |
| Manufacturer: | Stryker Instruments |
| Issue: | The manufacturing periodic testing to verify the consistency of the welds between the two sections of the housing was not conducted. Therefore, changes in the quality of the weld may not have been detected and may be less effective than intended. This could lead to the separation of the top section of the housing from the bottom section. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 17/10/2014 |