Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:17540
Brand Name:Stryker SmartLife Large Asceptic Housing
Model:7126120000
Affected:13027 to 14093
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Stryker Instruments
Issue:The manufacturing periodic testing to verify the consistency of the welds between the two sections of the housing was not conducted. Therefore, changes in the quality of the weld may not have been detected and may be less effective than intended. This could lead to the separation of the top section of the housing from the bottom section.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/10/2014