Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 17536 |
| Brand Name: | Siemens ADVIA Centaur and ADVIA Centaur XP, bulk fluid reserve assembly |
| Model: | Siemens Material Numbers: 10282187, 10483525, 10364516, and 10483526 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare Diagnostics Inc |
| Issue: | Manufacturer has determined that on some systems a slight cracking of the tubing (the tubing for the Wash1, acid, and base bottle reservoirs) at the connecting fitting on the reservoir connector assembly may occur due to mechanical stress.The cracking may result in a minor drip of fluid onto the reservoir and into the drip tray located underneath the reservoirs. The potential for minimal fluid leakage outside the system exists. The drip will cause the reservoir to empty faster than the system expects. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 17/10/2014 |