Detail of Recall

Type of Product:Device
Medsafe Reference:17536
Brand Name:Siemens ADVIA Centaur and ADVIA Centaur XP, bulk fluid reserve assembly
Model:Siemens Material Numbers: 10282187, 10483525, 10364516, and 10483526
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Manufacturer has determined that on some systems a slight cracking of the tubing (the tubing for the Wash1, acid, and base bottle reservoirs) at the connecting fitting on the reservoir connector assembly may occur due to mechanical stress.The cracking may result in a minor drip of fluid onto the reservoir and into the drip tray located underneath the reservoirs. The potential for minimal fluid leakage outside the system exists. The drip will cause the reservoir to empty faster than the system expects.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/10/2014