Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 17523 |
| Brand Name: | Siemens Artis Zeego Systems with a C-arm |
| Model: | C-arm |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens AG |
| Issue: | A cable inside the C-arm of systems affected may have been routed in an inappropriate manner and there is a potential that with increased wear a cable inside the C-arm may break. A cable break will result in restricted functionality and even system failure. The problem will not occur systematically but sporadically in the ongoing life time of a system. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 14/10/2014 |