Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:17523
Brand Name:Siemens Artis Zeego Systems with a C-arm
Model:C-arm
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Issue:A cable inside the C-arm of systems affected may have been routed in an inappropriate manner and there is a potential that with increased wear a cable inside the C-arm may break. A cable break will result in restricted functionality and even system failure.
The problem will not occur systematically but sporadically in the ongoing life time of a system.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/10/2014