Detail of Recall

Type of Product:Device
Medsafe Reference:17461
Brand Name:Siemens Advia Centaur Calibrator E
Model:Catalogue Numbers: 10309079 (2 pack), 10321075 (6 pack), 10335532 (6 pack-REF)
Affected:Kit lots ending in 38 (exp 29-Nov-2014), and 39 (15-May-2015)
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Siemens Healthcare Diagnostics has identified a positive bias for Calibrator E kit lots ending in 38 and 39 with the ADVIA Centaur Systems Cortisol and Progesterone assays compared to the respective master curves.
The observed positive bias with ADVIA Centaur Systems Cortisol and Progesterone assays with Calibrator E kit lots ending in 38 and 39 does not impact their clinical utility. The biases affect regions of the curve that are not clinically relevant in the investigation or monitoring of endocrine disorders or other uses.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/10/2014