Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:17436
Brand Name:Covidien Medi-Trace Cadence Defibrillator electrodes
Model:Model: 22660R
Affected:All lots
Software version:
Recalling Organisation:Covidien NZ Ltd
Level 3, Building 5, Central Park Corporate Centre
666 Great South Road
Penrose
AUCKLAND 1051
Contact Information:0508 489 264
Manufacturer:Covidien LLC
Issue:Covidien has received two reports where customers attempted to use a Covidien electrode with a Philips FR3 AED unit. These electrodes will not connect with Philips FR3 or FRx AED units, and the use of Covidien defibrillation electrodes with the Philips FR3 AED units, could result in a delay of therapy. The FRx AED unit requires the pads to be pre-connected, and will issue a continuous alarm chirp to alert the user that the proper pads are not connected to the unit prior to use.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:26/9/2014