Detail of Recall

Type of Product:Device
Medsafe Reference:17429
Brand Name:Siemens Dimension Vista Hemoglobin A1c (HbA1c) Flex Reagent Cartridge
Model:Catalogue Number: K3105A
Affected:Lots: 13294BA, 13308BA, 13329AA, 13350AA, 14006AA, 14041BA, 14069BB, 14083BA
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Siemens has confirmed that certain lots of the Dimension Vista® HbA1c Flex® reagent cartridge exhibit a positive bias averaging 0.4% [4.4 mmol/mol] Hemoglobin A1c units and occasionally up to 1.0% [11 mmol/mol] HbA1c units for patient samples when compared to the National Glycohemoglobin Standardization Program (NGSP). QC samples may exhibit a similar bias
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:25/9/2014