Detail of Recall

Type of Product:Device
Medsafe Reference:17380
Brand Name:Codman EDS3 External Drainage System with ventricular catheter
Model:Model: 82-1730
Affected:All lots with expiry on or before Aug-2017
Software version:
Recalling Organisation:Johnson & Johnson (NZ) Limited
13a Gabador Place
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:Codman & Shurtleff Inc
Brand Name:Codman EDS3 3CSF External Drainage System (no ventricular catheter)
Model:Model: 82-1731
Affected:All lots with expiry on or before Aug-2017
Software version:
Recalling Organisation:Johnson & Johnson (NZ) Limited
13a Gabador Place
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:Codman & Shurtleff Inc
Brand Name:Codman EDS3 Lumbar Drainage Catheter Kit II
Model:Model: 82-1738
Affected:All lots with expiry on or before Aug-2017
Software version:
Recalling Organisation:Johnson & Johnson (NZ) Limited
13a Gabador Place
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:Codman & Shurtleff Inc
Issue:The tubing within the system that drains CSF may leak or disconnect from the joints. Leakage and tube separations may result in over- or under-drainage of CSF from the ventricular system or introduction of air into the ventricular system (pneumocephalus). This may result in collapsed ventricles, subdural bleeding, or an inability to properly control elevated intracranial pressure. The tubing disconnection or leakage may also increase the risk of ventriculitis. If undetected or untreated each of these events may cause severe brain injury, which may lead to coma, stroke or death.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/9/2014