Detail of Recall

Type of Product:Device
Medsafe Reference:17370
Brand Name:Siemens ADVIA Centaur HBc Total Assay
Model:Catalogue Number: 07566733
Affected:Lots: 36217064, 36281064, 36449064, 37406064, 38177064
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Siemens Healthcare Diagnostics has confirmed increased assay imprecision and a positive bias (increase in index values) in the low end of the assay with the ADVIA Centaur® Systems HBc Total (HBcT) in selected kit lots. The increased imprecision and positive bias is observed with patient samples and negative controls and can result in an increase in false reactive results.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:16/9/2014