Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:17327
Brand Name:DePuy Charnley Pin Retractor & Handle Set
Model:Part Number - 962004000
Affected:Lot numbers: OSA-149328, OSA-161157, OSA-181610
Software version:
Recalling Organisation:Biomet New Zealand Ltd
Contact Information:0275 522 595
Manufacturer:DePuy International Ltd
Issue:Analysis of the pegs from the above affected lots shows the existence of microscopic cracks in the weld region. This could lead to decreased strength and a greater risk of instrument fracture under what would be normally acceptable conditions.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/9/2014
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /