Detail of Recall

Type of Product:Device
Medsafe Reference:17323
Brand Name:Covidien Puritan Bennett 840 Ventilator
Model:
Affected:
Software version:
Recalling Organisation:Covidien NZ Ltd
Level 3, Building 5, Central Park Corporate Centre
666 Great South Road
Penrose
AUCKLAND 1051
Contact Information:0508 489 264
Manufacturer:Covidien LLC
Issue:Manufacturer is providing new information for Operator's Manual. Customer reports that initiated these changes included the graphical user interface display screens (both upper and lower) becoming blank or erratic during patient use. This event interferes with the clinician's ability to view and/or change ventilator setting or clinician-set patient alarm parameters. The ventilator continues to provide respiratory support during these events.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/9/2014