| Type of Product: | Device |
| Medsafe Reference: | 17302 |
| Brand Name: | Draeger Babytherm Infant Warmers |
| Model: | Models: BT-8000 & BT-8010 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Draeger New Zealand Ltd Unit 4 24 Bishop Dunn Place East Tamaki Auckland 2013 |
| Contact Information: | 0800 559 186 |
| Manufacturer: | Draegerwerk AG & Co KGaA |
| Issue: | The manufacturer is aware of a case in which a side panel on a Babytherm 8010 did not fully slide to the bottom in the corner guide and was not completely locked. The side panel did not open during this incident and no patient was affected. Furthermore, if the inner 70mm panels were not always used and if a side panel has not been fully locked into position this may allow the baby's movements to open the side panel - it may be possible for a child to fall out. |
| Recall action type: | Alert |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 5/9/2014 |