Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:17302
Brand Name:Draeger Babytherm Infant Warmers
Model:Models: BT-8000 & BT-8010
Affected:
Software version:
Recalling Organisation:Draeger New Zealand Ltd
Unit 4
24 Bishop Dunn Place
East Tamaki
Auckland 2013
Contact Information:0800 559 186
Manufacturer:Draegerwerk AG & Co KGaA
Issue:The manufacturer is aware of a case in which a side panel on a Babytherm 8010 did not fully slide to the bottom in the corner guide and was not completely locked. The side panel did not open during this incident and no patient was affected.
Furthermore, if the inner 70mm panels were not always used and if a side panel has not been fully locked into position this may allow the baby's movements to open the side panel - it may be possible for a child to fall out.
Recall action type:Alert
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/9/2014