Detail of Recall

Type of Product:Device
Medsafe Reference:17290
Brand Name:Philips IntelliSpace PACS 4.4
Model:
Affected:
Software version:Ver 4.4
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Healthcare Informatics Inc
Issue:A problem has been detected in the Philips IntelliSpace PACS 4.4 software that, if it were to occur, could pose a risk for patients. Although the data is correct in IntelliSpace PACS, any changes to data may not be updated before Auto Export. If diagnosis is performed utilizing exported data on a 3rd party device the user may not be aware of any unchanged patient. If this were to occur diagnosis may be based on an incomplete set of images, or images that are associated with the wrong patient
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/9/2014