Detail of Recall

Type of Product:Device
Medsafe Reference:17251
Brand Name:GE Carescape Bx50 Patient Monitor
Model:Models B850, B650, B450
Affected:
Software version:PDM software version V2.2 or lower
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Medical Systems Information Technologies Inc
Issue:If the PDM is quickly disconnected and reconnected (a less than 10 second cycle) from a Carescape Bx50 monitor following a "Patient Discharge" or during active monitoring, the ECG waveform and its associated waveform parameters could be analyzed incorrectly.
The ECG heart rate (HR) parameter data, the ECG waveform analysis and other ECG parameter measurements are displayed at values that will be less than the actual patient's physiologic condition. As a result of the abnormal analysis of data, the alarms for serious conditions may not be triggered appropriately. The issue potentially affects heart rate, ECG ST segment analysis, 12 lead ECG analysis; ECG, SPO2, respiration and invasive blood pressure waveforms both locally and at CIC; and data displayed at CIC may show data drop out. Neither bedside nor CIC data can reliably represent the parameters being measured.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/9/2014