Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 17248 |
| Brand Name: | Cook Test Tube Heater, power supply unit |
| Model: | Part Number - K-APS-300 |
| Affected: | Lot Number: A928180, A928183, A928186 |
| Software version: | |
| Recalling Organisation: | Obex Medical Ltd Level 3, 109 Carlton Gore Road Newmarket Auckland 1023 |
| Contact Information: | 0800 656 539 |
| Manufacturer: | William A Cook Australia Pty Ltd |
| Issue: | The power supply of the devices are being replaced because recent testing of the combination of the test tube heater and power supply was found to exceed the prescribed electromagnetic compatibility (EMC) limits for radiated emissions. As the level of excess emissions are only slightly over the prescribed limits, and because the exposure to the radio frequency (RF) fields is far below internationally accepted safety levels, it is unlikely that there are any health risks to the user or patient posed by the non-conforming combination of the test tube heater (K-FTH-1012) and power supply (K-APS-300). It is also unlikely for the power supply to affect the operation of the test tube heater as this device is designed to have RF immunity greater than the radiation emitted by the power supply. |
| Recall action type: | Correction |
| Recall action: | Product to be exchanged |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 2/9/2014 |