Detail of Recall

Type of Product:Device
Medsafe Reference:17248
Brand Name:Cook Test Tube Heater, power supply unit
Model:Part Number - K-APS-300
Affected:Lot Number: A928180, A928183, A928186
Software version:
Recalling Organisation:Obex Medical Ltd
Level 3, 109 Carlton Gore Road
Newmarket
Auckland 1023
Contact Information:0800 656 539
Manufacturer:William A Cook Australia Pty Ltd
Issue:The power supply of the devices are being replaced because recent testing of the combination of the test tube heater and power supply was found to exceed the prescribed electromagnetic compatibility (EMC) limits for radiated emissions.
As the level of excess emissions are only slightly over the prescribed limits, and because the exposure to the radio frequency (RF) fields is far below internationally accepted safety levels, it is unlikely that there are any health risks to the user or patient posed by the non-conforming combination of the test tube heater (K-FTH-1012) and power supply (K-APS-300). It is also unlikely for the power supply to affect the operation of the test tube heater as this device is designed to have RF immunity greater than the radiation emitted by the power supply.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/9/2014