Detail of Recall

Type of Product:Device
Medsafe Reference:17222
Brand Name:Elekta Precise Treatment Table
Model:
Affected:
Software version:
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Elekta Limited (England)
Issue:Manufacturer identified that it is possible to get a positional error with the Precise Treatment Table. This can occur if the positional sensors are not correctly installed during corrective maintenance. The system has a software check for large positional errors which is intended to find faults, but it is possible to position the Treatment Table with errors greater than 5mm and no inhibits are displayed.
It is possible to cause a clinical mistreatment if the patient is not in the correct position.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/8/2014