Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 17222 |
| Brand Name: | Elekta Precise Treatment Table |
| Model: | |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Elekta Pty Limited - New Zealand Branch Suite 302, Geyser Building, 100 Parnell Road Parnell AUCKLAND 1052 |
| Contact Information: | 0273-544-793 |
| Manufacturer: | Elekta Limited (England) |
| Issue: | Manufacturer identified that it is possible to get a positional error with the Precise Treatment Table. This can occur if the positional sensors are not correctly installed during corrective maintenance. The system has a software check for large positional errors which is intended to find faults, but it is possible to position the Treatment Table with errors greater than 5mm and no inhibits are displayed. It is possible to cause a clinical mistreatment if the patient is not in the correct position. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 22/8/2014 |