Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 17217 |
| Brand Name: | Siemens CentraLink Data Management System |
| Model: | multiple material numbers |
| Affected: | |
| Software version: | below V14.0.4 |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare Diagnostics Inc |
| Issue: | The manufacturer has identified a task scheduler issue with the CentralLink Data Management System (version 14). The issue may cause the software to stop executing commands, including uploading validated results to the LIS. The issue is related to an internal software timer that overflows after 24 days. Automated rules triggered by task schedulers or by actions in the CentralLink user interface may also be affected. As each customer configuration is unique, in addition to the delay in uploading of validated results, the specific consequences of discontinued command execution will vary. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Healthcare Professional |
| Date Commenced: | 21/8/2014 |