Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:17215
Brand Name:Siemens ADVIA Centaur System/ ADVIA Centaur XP Immunoassay
Model:
Affected:Multiple material numbers
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Manufacturer has identified an issue with patient demographic information sent to the LIS from ADVIA Centar/ ADVIA Centaur XP Immunoassay systems. Under extremely rare circumstances patient demographic data from the previous order received from the LIS is merged with the next order.
This issue can occur when the LIS data buffer on the ADVIA Centair System becomes full and a particluar character if doun dn the last locations in the LIS data buffer
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/8/2014