Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 17182 |
| Brand Name: | Elekta Focal Radiation Therapy Treatment Planning Systems |
| Model: | |
| Affected: | |
| Software version: | Focal 4.7 and 4.8 when using Elekta or Siemens linacs |
| Recalling Organisation: | Elekta Pty Limited - New Zealand Branch Suite 302, Geyser Building, 100 Parnell Road Parnell AUCKLAND 1052 |
| Contact Information: | 0273-544-793 |
| Manufacturer: | Elekta Inc |
| Brand Name: | Elekta Monaco Radiation therapy treatment planning system |
| Model: | |
| Affected: | |
| Software version: | Monaco 3.20.00 and higher when using Elekta or Siemens linacs |
| Recalling Organisation: | Elekta Pty Limited - New Zealand Branch Suite 302, Geyser Building, 100 Parnell Road Parnell AUCKLAND 1052 |
| Contact Information: | 0273-544-793 |
| Manufacturer: | Elekta Inc |
| Issue: | When DICOM exporting an arc plan (VMAT, Dynamic Conformal or 3D Static Arcs) and using the Composite Field Sequencing (CFS) feature, the DICOM exported plan will not match the approved treatment plan. The problem occurs when the starting gantry angle of the second arc is exactly 180 degrees and the stop angle of first beam is greater than 180 degrees away from the start position of the second arc. CFS is designed to combine arcs into a single beam to streamline delivery. However, under these specific circumstances, the start and stop points for the arcs become corrupted. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 15/8/2014 |