Detail of Recall

Type of Product:Device
Medsafe Reference:17162
Brand Name:Steris 4085 series Surgical Table
Model:Models: 4085, 4085SS and 4085 Feather Weight Leg Sction
Affected:
Software version:
Recalling Organisation:Device Technologies New Zealand
47 Arrenway Drive
Albany
AUCKLAND 0632
Contact Information:09 913 2000
Manufacturer:Steris Corporation
Issue:STERIS has determined there is a potential that if the surgical table feet become unlocked from the floor during a procedure, the hand control may incorrectly indicate that the table feet are locked and still contacting the floor. The user may not realize that the table is unlocked and is able to continue articulating the table using the hand control. Should this occur, the table can be relocked by pressing the Floor Lock button on the table hand control.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/8/2014