Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:17158
Brand Name:Bard Arctic Sun Temperature Management System
Model:Catalogue Numbers: 5000-01, 748-00, 771-00
Affected:All lots
Software version:
Recalling Organisation:Pharmacy Retailing (NZ) Ltd t/a Healthcare Logistics
54 Carbine Road
Mt Wellington
AUCKLAND
Contact Information:09 969 0705
Manufacturer:Medivance Inc
Issue:Two adverse events have been reported to the manufacturer. In each instance, a temperature simulator key was connected to the patient temperature cable instead of the patient temperature probe. Due to the false patient temperature reading from the temperature simulator key, improper targeted temperature therapy was delivered to the patients by the Arctic Sun® 5000.
Recall action type:Alert
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/8/2014
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /