Detail of Recall

Type of Product:Device
Medsafe Reference:17152
Brand Name:Siemens Symbia T Series and Symbia Intevo camera system
Model:
Affected:Selected serial numbers
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Medical Solutions USA Inc
Issue:The manufacturer confirms a rubber component with unfavorable ageing properties. The mechanical function of this rubber damper, which is situated within the mounting assembly of the rotation motor, may potentially degrade over time and may impact stability.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/8/2014